Figure: 25 TAC §289.229(h)(2)(A)(i)

TABLE I.
System Leakage Limit Measurement Location
0-150 kVp (manufactured or installed
prior to March 1, 1989)
1 R (10 mGy) in 1 hr 1 meter (m) from source
0-150 kVp (manufactured on or after
March 1, 1989)
100 mR (1mGy) in 1 hr 1 m from source
151-499 kVp 1 R (10 mGy) in 1 hr 1 m from source